Remote Alarm Jack
Follow these steps to remove the remote alarm jack (). Reverse removal steps
to install.
1. Remove the top enclosure (see “Top Enclosure” on page 9-4).
2. Loosen the three captive screws and remove the screw holding the
PCB shroud, then remove the cover.
3. Remove the nut and washer from the remote alarm jack (Figure 9-58).
4. Disconnect the remote alarm jack harness from CN2 on the Main PCB
Figure 9-58: Remote Alarm Jack (Outside view)