Appendix A
A-14
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Oxygen Delivery
(Test 6, Step 1)
This step sets specific oxygen valve positions and oxygen flows to verify the
correct oxygen valve control and the accuracy of oxygen and exhalation flow
sensor readings.
Diagnostic codes
3102
,
3104
, and
3126
indicate that oxygen flow rates are
out of range. Diagnostic codes
3146
,
3147
, and
3148
indicate that the
oxygen valve stepper motor positions are out of range.
Red = air flow
Blue = O
2
flow
Black = no flow