Chapter 8
8-2
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Performance Verification
Removal/replacement: Power supply harness
Electrical safety
EST
Backup Battery and External Battery
Removal/replacement: Power switch
Electrical safety
EST
Removal/replacement: Rotary encoder
Electrical safety
EST
Removal/replacement: Blower
All
Removal/replacement: Sensor PCB
All
Removal/replacement: Main PCB
Electrical safety
EST
Alarm output signal, volume, and remote alarm
Backup Battery and External Battery
Removal/replacement: MMI PCB
Electrical safety
EST
Heated exhalation filter and power fail alarm
Backup Battery and External Battery
Removal/replacement: Blower motor
controller PCB
All
Removal/replacement: Motor controller(s)
All
Removal/replacement: CPU PCB
Electrical safety
EST
Removal/replacement: Digital PCB
Electrical safety
EST
Heated exhalation filter and power fail alarm
Removal/replacement: VGA PCB
Electrical safety
EST
Removal/replacement: Analog PCB
Electrical safety
EST
Air flow accuracy
Removal/replacement: Air flow sensor
Electrical safety
EST
Air flow accuracy
Gas volume accuracy
Removal/replacement: Air valve
All
Removal/replacement: Oxygen flow sensor
Electrical safety
EST
Oxygen flow accuracy
Gas volume accuracy
Oxygen accuracy
Removal/replacement: Oxygen valve
All
Service Performed
Required Tests
Table 8-2: Performance Verification Test Requirements