Chapter 9
9-70
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Component Removal/Installation
Figure 9-80: Oxygen Water Trap and Oxygen Regulator
Follow these steps to install the oxygen regulator assembly:
1. Fasten the plate and the oxygen regulator assembly to the bottom
enclosure using four 2.5-mm screws.
2. Connect the two black wires from the oxygen pressure switch to the J5
and J6 connectors on the sensor PCB (polarity is not important).
3. Use two tie wraps to fasten the two black wires to the bundle of cables
running across the top of the sensor PCB.
4. Install oxygen valve assembly (see “Oxygen Valve Assembly” on
page 9-67). Ensure that the oxygen valve is fully installed to the
regulator.
5. Install air valve assembly (see “Oxygen Regulator Assembly (with
Oxygen Pressure Switch)” on page 9-69).
6. Install air flow sensor (see “Air Flow Sensor” on page 9-65).
7. Install oxygen flow sensor (see “Oxygen Flow Sensor” on page 9-57).
Ensure that the oxygen flow sensor is fully installed to the oxygen valve
assembly.
8. Remove any residual Teflon tape from male and female threads
9. Apply 2¾ turns of Teflon tape to the thread of the inlet block.
Water trap bowl
Water trap bracket
Water trap
4-mm screws
2.5-mm
screws (4)
2.5-mm screws