Chapter 9
Component Removal/Installation
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
9-57
Figure 9-68: Filter Heater Tube Connection
Oxygen Flow Sensor
Follow these steps to remove the oxygen flow sensor (Figure 9-69):
1. Remove the top enclosure (see “Top Enclosure” on page 9-4).
2. Remove the enclosure brace.
3. Completely remove the 4-mm screws and clamp from the oxygen flow
sensor.
4. Cut the tie wrap holding cable O
2
F/S to the oxygen flow sensor, then
disconnect the cable.
5. Gently push the oxygen flow sensor and the aluminum sleeve toward
each other, then tilt the flow sensor up at the oxygen valve and lift out.
Filter
heater
assembly
Remove
silicone
tube from
tube nipple