Chapter 5
5-8
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Repair
Software Screen
To view the software screen, touch
Software
on the diagnostic screen (Figure
5-7). The software screen shows the:
•
Ventilator serial number.
•
Part number and version of the flash and one time programmable
(OTP) software.
•
Version of the air stepper valve, oxygen stepper valve, exhalation
stepper valve, and display.
•
Air, oxygen, and exhalation flow sensor part numbers.
•
Options installed, if any.
Figure 5-7: Diagnostic Mode Software Screen
Diagnostic Codes
The Esprit Ventilator generates diagnostic codes when the microprocessor
detects a fault during normal operation, or if a failure occurs during SST or
EST. Section 6 lists diagnostic codes, and recommended repair procedures for
each code.