Chapter 9
9-92
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Component Removal/Installation
4. Remove the two screws securing the original backlight inverter PCBA
to the bezel and two of the LCD bracket screws. Discard screws and
original backlight inverter PCBA (Figure 9-96).
Figure 9-96: Removing the Backlight Inverter
5. Remove the original 10.4-in. LCD and backlight inverter PCBA from
the GUI assembly.
6. If necessary, clean the inside surface of the touchframe using
Isopropyl Alcohol and a soft, lint-free cloth.
7. Connect the 2nd generation backlight inverter cable to the 2nd
generation backlight inverter PCBA.
8. Install the 2nd generation backlight inverter PCBA and secure with
two M2 x 0.45 x 6-mm screws (Figure 9-97).