Chapter 9
9-88
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Component Removal/Installation
Calibrating the Screen
1. Cycle ventilator power while holding down the
Exp. Hold
and left
Options
button.
2. The ventilator prompts you to calibrate the screen. Upon
acknowledgement, the software automatically detects and adjusts for
the 10.4-in. screen size.
3. After the ventilator confirms the screen calibration is complete, cycle
power to the ventilator and enter diagnostic mode.
4. Enter the software screen and confirm that the serial number is
correct.
5. Reset the altitude and compliance compensation settings in the user
configuration screen.
6. Perform an EST with the top enclosure removed. If the EST passes,
install the top enclosure (see “Top Enclosure” on page 9-4).
Enabling Options
Enable an option (see Chapter 7) only if the option upgrade is being installed
at this time.
Final Checkout
Complete the recommended tests in the performance verification procedure
(see section 8).
Replacing the GUI
(9.5-in. to 2nd
Generation 10.4-in.)
1. Disconnect power, unscrew and separate the ventilator top and bottom
halves. Set the halves side by side.
2. Disconnect all accessible wiring harnesses to the sensor, MMI, and
backlight inverter PCBs.
3. Remove the power supply shroud (see “Power Supply Fan/Shroud” on
4. Remove the hardware securing the power supply (see “Power Supply”
on page 9-10). Disconnect J2 and J3 harness connections to the
power supply. Position the power supply and remaining interconnect
wires so that they are out of the way.