Chapter 9
Component Removal/Installation
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
9-5
5. Carefully lift the top enclosure from the unit.
Figure 9-7: Top Enclosure
Sensor PCB
Follow these steps to remove the sensor PCB (Figure 9-8). Reverse removal
steps to install.
CAUTION:
When removing or installing the sensor PCB, avoid bumping the R76
potentiometer.
NOTE:
For a ventilator with a 9.5-in. display, the GUI and MMI PCB must be
installed before reinstalling the sensor PCB.
1. Remove the top enclosure (see “Top Enclosure” on page 9-4).
2. Remove all cables from the connectors on the sensor PCB:
NOTE: Tighten all enclosure
screws to 24 ± 4 in.-lbs.
M5 x 25 screw
M5 x 25 screw
M5 x 16 screw
Lockwashers
only on sides
Flat washers on
bottom only
M5 x 16 screws
Lock washers
Remove this cable:
From this connector on the
Sensor PCB:
SEN J1
J1
Black wires from oxygen pressure
switch*
J5 and J6
SEN J9
J9
SEN J2
J2