Chapter 9
9-6
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Component Removal/Installation
3. Disconnect the exhalation solenoid tube from pressure sensor U32,
and the inhalation solenoid tube from pressure sensor U33 on the
sensor PCB.
CAUTION:
To avoid damage to sensors, pry (don’t pull) tubes from sensors.
4. Unsnap the sensor PCB from the top two posts, then the bottom two
posts. Remove the sensor PCB.
Figure 9-8: Sensor PCB
SEN J7
J7
SEN J12
J12
SEN J10
J10
SEN J11
J11
SEN BRD J8
J8
SENSOR BRD J3 & J4*
J3 and J4
*Either wire can be connected to either terminal.
Remove this cable:
From this connector on the
Sensor PCB:
Do not bump R76
potentiometer (position on
original version of sensor PCB)
Pry (do not pull) tubes
from sensors