Chapter 9
Component Removal/Installation
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
9-21
3. Feed ribbon cables 6 and 7 through the remaining top enclosure slots,
then set the GUI assembly in position onto the top enclosure, ensuring
that all ribbon cables are untwisted.
4. Use the four 3-mm screws to reinstall the GUI to the top enclosure
Figure 9-23: GUI Cables
5. Readjust the intensity and volume potentiometers so that knobs can
turn freely. Do not overtighten screws.
6. Reposition and secure the power supply (see “Power Supply” on
7. Install the sensor PCB (see “Sensor PCB” on page 9-5).
8. Install the top enclosure (see “Top Enclosure” on page 9-4).
GUI Assembly
(Original 10.4-in.
GUI)
Follow these steps to remove the 10.4-in. GUI assembly (Figure 9-24, Figure
9-25). Reverse removal steps to install.
1. Unscrew the two 2.5-mm screws from the bottom corners of the GUI.
2. Remove the GUI knob cover, loosen the 11-mm nut, then remove the
knob.