Maintenance Procedures
QIAgility User Manual 11/2011
8-3
Note
: Do not wipe the rails. These rails support the pipetting
head and allow it to slide backwards and forwards easily.
Wiping the rails will remove the grease and make them more
susceptible to rust.
8.2
Periodic maintenance
8.2.1
Decontaminating the QIAgility
To remove DNA/RNA from the worktable and to eliminate
cross-contamination, the QIAgility must be routinely
decontaminated. Proceed as follows:
1.
Remove all loading blocks and the tip ejector chute from
the worktable. Wash these in detergent and rinse with
clean water.
2.
Soak these components in TriGene (Medichem
International; for more information, see www.medi-
mark.co.uk), or one of the tested cleaning agents listed
in Section 8.1.1 above, for 15–30 min.
3.
Thoroughly rinse the components in nucleic acid-free
water.
4.
Briefly rinse with absolute ethanol and dry with a soft
paper towel.
5.
Spray the instrument worktable with TriGene, or one of
the tested worktable cleaning agents listed above, and
allow to stand for 15–30 min. Wipe away with clean
paper towels.
6.
Spray the worktable the nucleic acid-free water and dry
with clean paper towels. Repeat this 3 times.
7.
Spray the worktable with TriGene, or one of the tested
worktable cleaning agents listed above, and allow to air
dry.
8.
Return all components to the worktable and close the
hood.
9.
If fitted, switch on the optional UV lamp for 15 min.
Summary of Contents for QIAgility
Page 10: ...Contents Contents 8 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 20: ...Safety Information 1 10 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 40: ...General Description 3 16 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 62: ...Installation Procedures 4 22 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 211: ...QIAgility Software QIAgility User Manual 11 2011 5 149...
Page 212: ...QIAgility Software 5 150 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 260: ...Maintenance Procedures 8 6 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 274: ...Troubleshooting 9 14 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 278: ...Glossary 10 4 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 284: ...Appendix A A 6 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 286: ...Appendix B B 2 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 298: ...Appendix C C 12 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 326: ...Appendix G G 4 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 328: ...Appendix H H 2 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 332: ...Index Index 4 QIAgility User Manual 11 2011 This page intentionally left blank...