QIAgility Software
QIAgility User Manual 11/2011
5-145
5.10.5
Checking the reaction setup
A reaction setup can be checked in 2 ways:
1.
Run a simulation of the reaction setup in “Virtual Mode”.
See Sections 5.1 and 5.2 for more information about
“Virtual Mode” operation.
2.
Place the cursor over a source well to check the
destination well(s) of the source well. The destination
wells will be highlighted. For instance, if the cursor is
placed over a master mix well, the wells on the reaction
plate that will contain that master mix are highlighted.
5.10.6
Required volumes for reaction setup
The software automatically calculates the volume of
standard, diluent, reagents, and master mix required for a
reaction setup. To view the volume of liquid required,
generate the pre-run report. This report will list the required
volumes. See Section 6.7.1 for more information about the
pre-run report.
Alternatively, position the cursor above a source well (i.e.,
diluent, reagent, master mix, or standard). A hover box will
appear that reports the volume required. This volume
includes an extra volume, to account for the dead volume,
that is dependent on the tube type.
5.11
Selecting the COM port
1.
Select “Options/Robot Setup/Select COM port”.
2.
The “Configure Serial Interface” dialog box will open.
Summary of Contents for QIAgility
Page 10: ...Contents Contents 8 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 20: ...Safety Information 1 10 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 40: ...General Description 3 16 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 62: ...Installation Procedures 4 22 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 211: ...QIAgility Software QIAgility User Manual 11 2011 5 149...
Page 212: ...QIAgility Software 5 150 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 260: ...Maintenance Procedures 8 6 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 274: ...Troubleshooting 9 14 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 278: ...Glossary 10 4 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 284: ...Appendix A A 6 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 286: ...Appendix B B 2 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 298: ...Appendix C C 12 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 326: ...Appendix G G 4 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 328: ...Appendix H H 2 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 332: ...Index Index 4 QIAgility User Manual 11 2011 This page intentionally left blank...