QIAgility Software
QIAgility User Manual 11/2011
5-89
Preserve source
well indices?
This function is visible only when enabled
in the “Advanced Settings” menu in the
“Options” menu. When selected, the
reaction constructed for each sample is
placed into the same relative location
from the start of the reaction plate as the
sample from the start of the sample plate,
determined according to pipetting order.
Mix on
ejection?
Instructs the instrument to perform a mix
cycle upon ejecting the sample and
defines the number of mix cycles to
perform. Mixing is only performed when
the final reaction component is added,
and only when there is more than 1
component in the reaction.
Use
normalization?
Instructs the instrument to normalize
sample concentrations.
Auto Adjust
small Volumes
Instructs the instrument to round up
sample volumes that are less than the
minimum volume that can be pipetted
(i.e., 1 µl), to the minimum volume (i.e.,
1 µl). Where the operation would require
less than the pipettable volume of diluent,
no diluent is transferred, only sample is
moved. In all such cases, the final volume
is preserved, but the desired final
concentration is not achieved.
Normalize to
conc.
Defines the concentration to which
samples will be normalized. This excludes
the contribution that other reagents and
master mix make to the reaction.
Setting up a reaction list
Note
: If the total number of reactions configured on one
plate exceeds the number of reaction wells a warning will
Summary of Contents for QIAgility
Page 10: ...Contents Contents 8 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 20: ...Safety Information 1 10 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 40: ...General Description 3 16 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 62: ...Installation Procedures 4 22 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 211: ...QIAgility Software QIAgility User Manual 11 2011 5 149...
Page 212: ...QIAgility Software 5 150 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 260: ...Maintenance Procedures 8 6 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 274: ...Troubleshooting 9 14 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 278: ...Glossary 10 4 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 284: ...Appendix A A 6 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 286: ...Appendix B B 2 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 298: ...Appendix C C 12 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 326: ...Appendix G G 4 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 328: ...Appendix H H 2 QIAgility User Manual 11 2011 This page intentionally left blank...
Page 332: ...Index Index 4 QIAgility User Manual 11 2011 This page intentionally left blank...