High-level ITS Installation and Configuration Procedure
IntelliVue Telemetry System Infrastructure Installation and Service Guide
3
-
3
e) Verify that the AP moves to Registered APs list.
f) Disconnect and then reconnect the AP to the network.
g) Verify that the AP is associated to the correct APC and has its IP address set
correctly.
10. Change the names of installed APs and Remote Antennas from their default
values to meaningful, user-friendly names. (
page 3-40
)
11. Run the Philips Upgrade Wizard again to verify that the APCs and APs have the
same firmware revision and proper configuration settings. (
page 3-41
)
12. Export the ITS configuration to a disk file for archive purposes. (
page 3-49
)
13. Restore your service PC to its original settings. (
page 3-49
)
14. Enter the ITS infrastructure devices into the Database Server/Standalone
IntelliVue Information Center configuration wizard, and then bring the ITS
bedside monitors and patient-worn transceivers online. (
page 3-50
)