Step 11. Run the Philips Upgrade Wizard Again.
IntelliVue Telemetry System Infrastructure Installation and Service Guide
3
-
41
Standard APs will list Remote Antennas as Not Present. Core APs will provide
general status of connected Remote Antennas.
4. Click SAVE and then verify the AP Configuration settings.
Step 11. Run the Philips Upgrade Wizard Again.
Run the Philips Upgrade Wizard again to check and verify final APC and AP
configurations, and fix any errors as needed. If any configuration modifications are
made, run the Philips Upgrade Wizard again.
Caution
Close the APC web interface before running the Philips Upgrade Wizard. Do not add
components to or remove components from the IntelliVue Telemetry System while
the wizard is running.
Follow these steps to upgrade ITS Access Point Controllers and Access Points using
the Philips Upgrade Wizard:
Philips Upgrade Wizard:
1. Run the Philips Upgrade Wizard by double-clicking the file:
C:\TelemetryUpgrader\upgrader.exe
The Upgrade Wizard splash screen (Figure 3-25) appears.
2. Click Next> to continue.
The APC/AP firmware selection screen (Figure 3-18) appears.
3. Complete the settings on the APC/AP firmware selection screen:
a) Click ... and then browse to the directory (e.g.,
C:\TelemetryUpgrader\AP
and APC Images\)
in which you have installed the APC and AP firmware
files.
b) Specify what type of APs you have installed by marking the appropriate radio
button, 1.4 GHz Smart Hopping or 2.4 GHz Smart Hopping.
Figure 3-25: Upgrade Wizard Splash Screen
Summary of Contents for IntelliVue Telemetry System Infrastructure
Page 8: ...viii Contents ...
Page 14: ...xiv About This Guide ...
Page 100: ...3 22 Chapter 3 Installing and Configuring the ITS Figure 3 10 APC Filter Configuration Screen ...
Page 162: ...5 12 Chapter 5 Troubleshooting and Testing ...
Page 168: ...A 6 Chapter A Installing Multiple ITSs at a Single Hospital Site ...
Page 180: ...Index 8 Index ...