D-6
PR Overrange
(Mindray,
Masimo,
Nellcor)
No Low
The measured PR
value exceeds the
measurement
range.
Contact your service
personnel.
‘‘*’’ means the alarm level is user-adjustable.
D.2.3
Temperature Alarm Messages
Alarm
Message
Clearable?
(Yes/No)
Alarm
Level
Cause
Solution
Temp Init
Err
Yes High
An error occurred to
the Temp module, or
there is a problem
with the
communications
between the module
and the monitor, or
Temp calibration error.
Restart the monitor. If the
error remains, contact
your service personnel.
Temp
Comm Err
No High
An error occurred to
the Temp module, or
there is a problem
with the
communications
between the module
and the monitor.
Restart the monitor. If the
error remains, contact our
service personnel.
Temp Alarm
Limit Err
No High
The alarm limit of
Temp is changed
inadvertently.
Contact your service
personnel.
Warmup
Timed Out
Yes Med
The initial probe
temperature in
measurement is too
high.
Cool the probe and retry.
Warming
Resistor Err
No Med
The thermal resistor
on the temperature
probe has an error
(Can not work
properly).
Replace temperature
probe.
Ambient
Temp
Overrange
Yes Med
The environmental
temperature is out the
range of the monitor's
measurement.
Change an environment
and retry.
Temp
Voltage Err
Yes Med
The voltage is too high
or too low.
Check the power supply.
Summary of Contents for VS-900
Page 1: ...VS 900 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 14: ...8 FOR YOUR NOTES ...
Page 26: ...2 6 2 3 4 Bottom View 1 Battery compartment door 2 Hole for installing a support 1 2 ...
Page 42: ...3 10 FOR YOUR NOTES ...
Page 68: ...6 12 FOR YOUR NOTES ...
Page 76: ...7 8 FOR YOUR NOTES ...
Page 78: ...8 2 FOR YOUR NOTES ...
Page 88: ...9 10 FOR YOUR NOTES ...
Page 114: ...12 8 FOR YOUR NOTES ...
Page 139: ...16 11 2 Select Modify Password 3 Enter new password and then select Ok ...
Page 140: ...16 12 FOR YOUR NOTES ...
Page 166: ...B 8 FOR YOUR NOTES ...
Page 191: ......
Page 192: ...P N 046 004883 00 7 0 ...