A-12
CO
2
resolution
1 mmHg
Sample flow rate
50 ml/min
Sample flow rate
tolerance
15% or 15 ml/min, whichever is greater.
Start-up time
< 90 s
Response time
5 s @ 50ml/min
Rise time
< 200 ms @ 50ml/min
awRR measurement range
0 to 150 rpm
awRR measurement
precision
<60 bpm: ±1
60 to 150 bpm: ±2
awRR resolution
1 rpm
Effect of interference gases on CO2 measurement
Gas
Concentration (%)
Quantitative
effect*
N
2
O ≤
60
±1 mmHg
Hal ≤
4
Sev ≤
5
Iso ≤
5
Enf ≤
5
Des
≤ 15
±2 mmHg
*: means an extra error should be added in case of gas interference when CO2 measurements
are performed between 0 to 40mmHg. Inaccuracy specifications are affected by the breath
rate and I:E change. The end-tidal gas reading is within specification for breath rate below
15BPM and I:E ratio smaller than 1:1 relative to the gas readings without breath.
Alarm limit
Range
Step
EtCO
2
High
(low limit + 2) to 99 mmHg
1 mmHg
EtCO
2
Low
1 to (high limit - 2) mmHg
FiCO
2
High
1 to 99 mmHg
awRR High
Adult, pediatric:
Neonate:
(low limit + 2) to 100 rpm
(low limit + 2) to 150 rpm
1 rpm
awRR Low
0 to (high limit - 2) rpm
Summary of Contents for VS-900
Page 1: ...VS 900 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 14: ...8 FOR YOUR NOTES ...
Page 26: ...2 6 2 3 4 Bottom View 1 Battery compartment door 2 Hole for installing a support 1 2 ...
Page 42: ...3 10 FOR YOUR NOTES ...
Page 68: ...6 12 FOR YOUR NOTES ...
Page 76: ...7 8 FOR YOUR NOTES ...
Page 78: ...8 2 FOR YOUR NOTES ...
Page 88: ...9 10 FOR YOUR NOTES ...
Page 114: ...12 8 FOR YOUR NOTES ...
Page 139: ...16 11 2 Select Modify Password 3 Enter new password and then select Ok ...
Page 140: ...16 12 FOR YOUR NOTES ...
Page 166: ...B 8 FOR YOUR NOTES ...
Page 191: ......
Page 192: ...P N 046 004883 00 7 0 ...