16-10
5.
Turn on and adjust the relief valve to make the flowmeter reads within 10-50ml/min
and keeps stable as well.
6.
In the [
Calibrate CO2
] menu, enter the vented CO
2
concentration in the [
CO2
] field.
7.
In the [
Calibrate CO2
] menu, the measured CO
2
concentration is displayed. After the
measured CO
2
concentration becomes stable, select [
Calibrate
] to calibrate the CO
2
module.
If the calibration is finished successfully, the message [
Calibration Completed!
] is
displayed in the [
Calibrate CO2
] menu. If the calibration failed, the message [
Calibration
Failed!
] is displayed. In this case, perform another calibration.
16.7
Calibrating the Touchscreen
1. Select
[
Main
]
→
[
Maintenance>>
]
→
[
Calibrate Touchscreen
]. The
symbol will
appear at different positions of the screen.
2.
Select, in turn the central point of the
symbol. After the calibration is
completed, the message [
Screen Calibration Completed
!
] is displayed.
3. Select
[
Ok
] to confirm the completion of the calibration.
16.8
Formatting the Storage Card
The monitor is configured with an SD card for saving data. To format the storage card:
1. Select
[
Main
]
→
[
Maintenance >>
]
→
[
User Settings >>
]
→
Enter the required
password
→
Select [
Ok
] to access the [
User Settings]
menu.
2. Select
[
Format Storage Card
], and then select [
Ok
] in the pop-up dialog.
16.9
Modifying Password
1. Select
[
Main
]
→
[
Maintenance >>
]
→
[
User Settings >>
]
→
Enter the required
password
→
Select [
Ok
] to access the [
User Settings
] menu.
Flowmeter
Tubing
T-shape connector
Monitor
Relief valve
Gas cylinder
Summary of Contents for VS-900
Page 1: ...VS 900 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 14: ...8 FOR YOUR NOTES ...
Page 26: ...2 6 2 3 4 Bottom View 1 Battery compartment door 2 Hole for installing a support 1 2 ...
Page 42: ...3 10 FOR YOUR NOTES ...
Page 68: ...6 12 FOR YOUR NOTES ...
Page 76: ...7 8 FOR YOUR NOTES ...
Page 78: ...8 2 FOR YOUR NOTES ...
Page 88: ...9 10 FOR YOUR NOTES ...
Page 114: ...12 8 FOR YOUR NOTES ...
Page 139: ...16 11 2 Select Modify Password 3 Enter new password and then select Ok ...
Page 140: ...16 12 FOR YOUR NOTES ...
Page 166: ...B 8 FOR YOUR NOTES ...
Page 191: ......
Page 192: ...P N 046 004883 00 7 0 ...