9-3
9.3
Measurement Limitations
Measurements are impossible with heart rate extremes of less than 40 bpm or greater
than 240bpm, or if the patient is on a heart-lung machine.
The measurement may be inaccurate or impossible:
If arterial pressure pulses are hard to detect
In the presence of excessive and continuous patient movement such as
shivering or convulsions
During certain cardiac arrhythmias
For pregnant or pre-eclamptic patients
Rapid blood pressure changes
Severe shock or hypothermia that reduces blood flow to the peripheries
Obesity, where a thick layer of fat surrounding a limb dampens the oscillations
coming from the artery
9.4
NIBP Measurement Mode
There are the following modes of measuring NIBP:
Manual: measurement on demand.
Auto: continually repeated measurements at set intervals.
STAT: continually rapid series of measurements over a five minute period, then
return to the previous mode.
Program: continually automatic measurement at set durations and intervals.
To set NIBP measurement mode, select NIBP parameter area to access [
NIBP Setup
]
menu:
Select [
Interval
] to select manual mode or auto NIBP measurement interval.
Select [
NIBP STAT
] to start a continuous NIBP measurement.
Summary of Contents for VS-900
Page 1: ...VS 900 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 14: ...8 FOR YOUR NOTES ...
Page 26: ...2 6 2 3 4 Bottom View 1 Battery compartment door 2 Hole for installing a support 1 2 ...
Page 42: ...3 10 FOR YOUR NOTES ...
Page 68: ...6 12 FOR YOUR NOTES ...
Page 76: ...7 8 FOR YOUR NOTES ...
Page 78: ...8 2 FOR YOUR NOTES ...
Page 88: ...9 10 FOR YOUR NOTES ...
Page 114: ...12 8 FOR YOUR NOTES ...
Page 139: ...16 11 2 Select Modify Password 3 Enter new password and then select Ok ...
Page 140: ...16 12 FOR YOUR NOTES ...
Page 166: ...B 8 FOR YOUR NOTES ...
Page 191: ......
Page 192: ...P N 046 004883 00 7 0 ...