15-4
2. Install
the
battery.
3.
Connect the monitor to AC power. Allow the battery to be charged uninterrupted till
the battery is full and the battery indicator is off.
4.
Remove AC power and allow the monitor to run from the battery until it shuts off.
The operating time of a battery reflects their performance directly. If the operating time of
a battery is noticeably shorter than that stated in the specifications, contact your service
personnel.
NOTE
The battery might be damaged or malfunctioned if its operating time is too
short after being fully charged. The operating time depends on the
configuration and operation. For example, measuring NIBP more frequently will
also shorten the operating time.
When a battery has visual signs of damage, or no longer holds a charge, it
should be replaced. Remove the old battery from the monitor and recycle it
properly.
15.6
Recycling a Battery
When a battery has visual signs of damage, or no longer holds a charge, it should be
replaced. Remove the old battery from the monitor and recycle it properly. To dispose of
the batteries, follow local laws for proper disposal.
WARNING
Do not disassemble batteries, or dispose of them in fire, or cause them to short
circuit. They may ignite, explode, leak or heat up, causing personal injury.
Summary of Contents for VS-900
Page 1: ...VS 900 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 14: ...8 FOR YOUR NOTES ...
Page 26: ...2 6 2 3 4 Bottom View 1 Battery compartment door 2 Hole for installing a support 1 2 ...
Page 42: ...3 10 FOR YOUR NOTES ...
Page 68: ...6 12 FOR YOUR NOTES ...
Page 76: ...7 8 FOR YOUR NOTES ...
Page 78: ...8 2 FOR YOUR NOTES ...
Page 88: ...9 10 FOR YOUR NOTES ...
Page 114: ...12 8 FOR YOUR NOTES ...
Page 139: ...16 11 2 Select Modify Password 3 Enter new password and then select Ok ...
Page 140: ...16 12 FOR YOUR NOTES ...
Page 166: ...B 8 FOR YOUR NOTES ...
Page 191: ......
Page 192: ...P N 046 004883 00 7 0 ...