6-9
The icon
appears in the alarm symbol area.
The parameter numeric and alarm limits still flash.
The indication of alarm lamp for the physiological alarm depends on the alarm light
setting.
When [
Alarm Light on Alarm Reset
] is set to [
On
], the alarm lamp remains
flashing.
When [
Alarm Light on Alarm Reset
] is set to [
Off
], the alarm lamp stops flashing.
To set [
Alarm Light on Alarm Reset
]:
1. Select
[
Main
]
→
[
Maintenance >>
]
→
[
User Settings >>
]
→
enter the required
password
→
Select [
Ok
] to access [
User Settings
] menu.
2. Select
[
Alarm Setup >>
] to enter the [
Alarm Setup
] menu.
3. Select
[
Alarm Light on Alarm Reset
], and toggle between [
On
] and [
Off
].
The default setting for [
Alarm Light on Alarm Reset
] is [
On
].
Technical alarms give different alarm indicators when the alarm system is reset:
For some technical alarms, including the NIBP-related alarms, a
√
appears
before the alarm message and
appears in the alarm symbol area,
indicating that the alarm is acknowledged. The indication of the alarm lamp
depends on the alarm light setting.
Some technical alarms are changed to the prompt messages.
Some technical alarms are cleared. The monitor gives no alarm indications.
For details about the indications of technical alarms when the alarm system is reset, refer
to
D.2 Technical Alarm Messages
.
6.9
Setting the Reminder Tone
In audio off status, the monitor can issue a periodical reminder tone. The interval of the
reminder tone is 1 minute. You can switch on or off the reminder tone.
1. Select
[
Main
]
→
[
Maintenance >>
]
→
[
User Settings >>
]
→
Enter required password
→
Select [
Ok
] and then access [
User Settings
] menu.
2. Select
[
Alarm Setup >>
] to access [
Alarm Setup
] menu.
3. Set
[
Reminder Tone
] to [
On
] or [
Off
].
Summary of Contents for VS-900
Page 1: ...VS 900 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 14: ...8 FOR YOUR NOTES ...
Page 26: ...2 6 2 3 4 Bottom View 1 Battery compartment door 2 Hole for installing a support 1 2 ...
Page 42: ...3 10 FOR YOUR NOTES ...
Page 68: ...6 12 FOR YOUR NOTES ...
Page 76: ...7 8 FOR YOUR NOTES ...
Page 78: ...8 2 FOR YOUR NOTES ...
Page 88: ...9 10 FOR YOUR NOTES ...
Page 114: ...12 8 FOR YOUR NOTES ...
Page 139: ...16 11 2 Select Modify Password 3 Enter new password and then select Ok ...
Page 140: ...16 12 FOR YOUR NOTES ...
Page 166: ...B 8 FOR YOUR NOTES ...
Page 191: ......
Page 192: ...P N 046 004883 00 7 0 ...