2-5
2.3.3
Rear View
(1) Temperature
probe
well
The temperature module varies in appearance. The above figure shows the main unit
with the SmarTemp
TM
module.
(2) Connector for temperature probe
The temperature module varies in appearance. The above figure shows the main unit
with the SmarTemp
TM
module.
(3) Network connector: It is a standard RJ45 connector used to communicate with
external devices, such as central monitoring system, Gateway, or upgrade the system.
(4) Multi-function connector: connects to the hospital's nurse call system, or connects
external devices through DIAP protocol.
(5) USB connector: connects to barcode scanner or USB disk.
(6) AC power input
(7) Equipotential grounding terminal
When the equipment and other devices are to be used together, their equipotential
grounding terminals should be connected together to eliminate the potential
difference between them.
(1)
(2)
(3)
(4)
(5)
(6)
(7)