A-2
Item
Operating conditions
Storage conditions
Temperature
16ºC to 33 ºC (60.8ºF to 91.4 ºF)
-25 to 55ºC (-13ºF to 131ºF)
Relative humidity
(non-condensing)
10% to 95%
up to 95%
WARNING
The equipment may not meet the performance specifications if stored or used
outside the specified temperature and humidity ranges.
NOTE
The environmental specifications of unspecified parameters are the same as
those of the main unit.
A.3
Power Supply Specifications
AC Power
Line voltage
100 to 240 VAC ~ (
±
10%)
Current
0.9 to 0.5A
Frequency
50/60 Hz (
±
3Hz)
Fuse T2AL-250V
Battery
Battery (Standard)
Battery Type
LI13I001A, Chargeable Lithium-Ion
Voltage 11.1
VDC
Capacity
2600 mAh
Run time
At least 3 hours when powered by a new fully-charged
battery at 25°C ±5°C with SpO
2
cable connected, and auto
NIBP measurements at an interval of 15 minutes.
Charge time
Monitor power off: less than 3 hours to 90%; less than 4 hours
to 100%.
Monitor power on: less than 6 hours to 90%; less than 7.5
hours to 100%.
Summary of Contents for VS-900
Page 1: ...VS 900 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 14: ...8 FOR YOUR NOTES ...
Page 26: ...2 6 2 3 4 Bottom View 1 Battery compartment door 2 Hole for installing a support 1 2 ...
Page 42: ...3 10 FOR YOUR NOTES ...
Page 68: ...6 12 FOR YOUR NOTES ...
Page 76: ...7 8 FOR YOUR NOTES ...
Page 78: ...8 2 FOR YOUR NOTES ...
Page 88: ...9 10 FOR YOUR NOTES ...
Page 114: ...12 8 FOR YOUR NOTES ...
Page 139: ...16 11 2 Select Modify Password 3 Enter new password and then select Ok ...
Page 140: ...16 12 FOR YOUR NOTES ...
Page 166: ...B 8 FOR YOUR NOTES ...
Page 191: ......
Page 192: ...P N 046 004883 00 7 0 ...