A-7
0% to 69%: Not specified.
Refreshing rate
1 s
Response time
≤ 30 s (PI > 0.3, no disturbance, SpO2 value sudden change
within 70% - 100%)
* When the SpO
2
sensor is applied for neonatal patients as indicated, the specified accuracy
range is increased by ±1%, to compensate for the theoretical effect on oximeter
measurements of fetal hemoglobin in neonatal blood.
Information of the Test Subjects of the Clinical Study Report
Skin color
Gender
Number
Age (years)
Health
Black Male 1
28.2±9.19
Healthy
Female 1
Yellow Male 3
Female 9
A.6.2
PR
Module
PR High Limit
PR Low Limit
Step (bpm)
Mindray SpO
2
Module
(low limit +2) to 254
20 to (high limit-2)
1
Masimo SpO
2
Module
(low limit +2) to 240
25 to (high limit-2)
Nellcor SpO
2
Module
(low limit +2) to 300
20 to (high limit-2)
NIBP Module
(low limit +2) to 240
40 to (high limit-2)
PR from Mindray SpO
2
Module
Measurement range
20 to 254 bpm
Resolution 1
bpm
Response time
≤ 30 s (PI > 0.3, no disturbance, PR value sudden change within
25 --- 250 bpm)
Accuracy
±3 bpm (without motion)
Refreshing rate
1 s
PR from Masimo SpO
2
Module
Measurement range
25 to 240 bpm
Resolution 1
bpm
Summary of Contents for VS-900
Page 1: ...VS 900 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 14: ...8 FOR YOUR NOTES ...
Page 26: ...2 6 2 3 4 Bottom View 1 Battery compartment door 2 Hole for installing a support 1 2 ...
Page 42: ...3 10 FOR YOUR NOTES ...
Page 68: ...6 12 FOR YOUR NOTES ...
Page 76: ...7 8 FOR YOUR NOTES ...
Page 78: ...8 2 FOR YOUR NOTES ...
Page 88: ...9 10 FOR YOUR NOTES ...
Page 114: ...12 8 FOR YOUR NOTES ...
Page 139: ...16 11 2 Select Modify Password 3 Enter new password and then select Ok ...
Page 140: ...16 12 FOR YOUR NOTES ...
Page 166: ...B 8 FOR YOUR NOTES ...
Page 191: ......
Page 192: ...P N 046 004883 00 7 0 ...