10-7
4.
Place the probe cover package on the loader, and take away the slip to let the probe
covers fall into the loader.
NOTE
It is recommended to quickly take away the slip to prevent the probe covers
from turning over in the loader.
5.
Push down the cap. Make sure that the cap is aligned with the notch on the loader.
Then turn clockwise to lock the cap.
6.
Push back the loader into the dock.
7.
Replace the ear thermometer to the dock.
10.2.1.3
Taking an Ear Temperature
WARNING
Do not reuse the probe cover. Always use a new, undamaged and clean probe
cover for a measurement.
Use only the specified probe cover.
To take an ear measurement and transfer it to the monitor:
1.
Power on the monitor.
2.
Remove the ear thermometer from the dock.
3.
Slide the lever to the end. A probe cover comes out when sliding the lever.
4. Release
the
lever.
5.
Firmly press the probe tip into the probe cover until you hear a click.