A-8
Response time
≤ 20 s (PR value sudden change within 25 --- 240 bpm)
Accuracy
±3 bpm (without motion)
±5 bpm (with motion)
Refreshing rate
1 s
PR from Nellcor SpO
2
Module
Measurement range
20 to 300 bpm
Resolution 1
bpm
Response time
≤ 30 s (PI > 0.3, no disturbance, PR value sudden change within
25 --- 250 bpm)
Accuracy
20 to 250 bpm: ±3 bpm
251 to 300 bpm, not specified
Refreshing rate
1 s
PR from N
IBP
Module
Measurement range
40 to 240 bpm
Resolution 1
bpm
Accuracy
±3bpm or ±3%, whichever is greater
A.6.3
NIBP
Standards
Meet standards of IEC 80601-2-30, ISO 81060-2
Technique Oscillometry
Mode of operation
Manual, Auto, STAT and Program
Auto mode repetition
intervals
1, 2, 2.5, 3, 5, 10, 15, 20, 30, 60, 90 or 120 min
STAT mode cycle time
5 min
Max measurement time
Adult, pediatric:
180 s
Neonate:
90 s
Heart rate range
40 to 240 bpm
Measurement ranges
(mmHg)
Adult
Pediatric
Neonate
Systolic:
40 to 270
40 to 200
40 to 135
Diastolic:
10 to 210
10 to 150
10 to 100
Mean:
20 to 230
20 to 165
20 to 110
Summary of Contents for VS-900
Page 1: ...VS 900 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 14: ...8 FOR YOUR NOTES ...
Page 26: ...2 6 2 3 4 Bottom View 1 Battery compartment door 2 Hole for installing a support 1 2 ...
Page 42: ...3 10 FOR YOUR NOTES ...
Page 68: ...6 12 FOR YOUR NOTES ...
Page 76: ...7 8 FOR YOUR NOTES ...
Page 78: ...8 2 FOR YOUR NOTES ...
Page 88: ...9 10 FOR YOUR NOTES ...
Page 114: ...12 8 FOR YOUR NOTES ...
Page 139: ...16 11 2 Select Modify Password 3 Enter new password and then select Ok ...
Page 140: ...16 12 FOR YOUR NOTES ...
Page 166: ...B 8 FOR YOUR NOTES ...
Page 191: ......
Page 192: ...P N 046 004883 00 7 0 ...