11-2
Measurement accuracy may be affected by the breath rate and inspiration/expiration
ratio (I: E ratio) as follow:
EtCO
2
value is within specification for breath rate ≤ 60 bpm and I/E ratio ≤ 1:1;
EtCO
2
value is within specification for breath rate ≤ 30 bpm and I/E ratio ≤ 2:1.
Measurement accuracy is unspecified for breath rate larger than 60 bpm.
11.4
CO
2
Display
The CO
2
parameter and waveform areas provide FiCO
2
measurement, EtCO
2
measurement,
awRR measurement, and a CO
2
waveform.
(1)
End tidal CO
2
(EtCO
2
)
(2)
Fraction of inspired CO
2
(FiCO
2
)
(3)
Airway respiration rate (RR)
(4) CO
2
waveform
(5) Scale
11.5
Measuring CO
2
CAUTION
Eliminate the exhausted gas before performing the measurement.
Connect an exhaust tube to the gas outlet connector of the monitor to vent the
calibration gases to a scavenging system.
Check that the alarm limit settings are appropriate before taking measurement.
To measure CO
2
, follow this procedure:
1.
Select the appropriate sampling line according to the patient category.
2.
Connect the sampling line to the gas sampling port of the mini water trap by
clockwise turning.
(1)
(2)
(4)
5
(3)
Summary of Contents for VS-900
Page 1: ...VS 900 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 14: ...8 FOR YOUR NOTES ...
Page 26: ...2 6 2 3 4 Bottom View 1 Battery compartment door 2 Hole for installing a support 1 2 ...
Page 42: ...3 10 FOR YOUR NOTES ...
Page 68: ...6 12 FOR YOUR NOTES ...
Page 76: ...7 8 FOR YOUR NOTES ...
Page 78: ...8 2 FOR YOUR NOTES ...
Page 88: ...9 10 FOR YOUR NOTES ...
Page 114: ...12 8 FOR YOUR NOTES ...
Page 139: ...16 11 2 Select Modify Password 3 Enter new password and then select Ok ...
Page 140: ...16 12 FOR YOUR NOTES ...
Page 166: ...B 8 FOR YOUR NOTES ...
Page 191: ......
Page 192: ...P N 046 004883 00 7 0 ...