D-4
Alarm
Message
Clearable?
(Yes/No)
Alarm
Level
Cause
Solution
NIBP Cuff
Type Wrong
Yes Low
The cuff type
applied mismatches
the patient category.
Verify the patient category
and replace the cuff.
NIBP Illegally
Reset
Yes Low
An illegal reset
occurred during
NIBP measurement.
Check if the airway is
occluded.
VeniPuncture
timeout
Yes Low
System deflates the
cuff after a certain
time.
/
D.2.2
SpO
2
Alarm Messages
Alarm
Message
Clearable?
(Yes/No)
Alarm
Level
Cause
Solution
SpO
2
Sensor
Off
(Mindray,
Masimo,
Nellcor)
Yes
Low*(Mon
itor mode)
Off* (Spot
Check
mode)
The SpO
2
sensor
has become
detached from the
patient or the
module, or there is
a fault with the
SpO
2
sensor, or an
unspecified SpO
2
sensor has been
used.
Check the sensor
application site and the
sensor type, and make
sure if the sensor is
damaged. Reconnect the
sensor or use a new
sensor.
SpO
2
Sensor
Fault
(Mindray,
Masimo)
No Low
SpO
2
No
Sensor
(Mindray,
Masimo,
Nellcor)
Yes Low
SpO
2
Unknown
Sensor
(Masimo)
No Low
SpO
2
Too
Much Light
(Mindray,
Masimo)
No Low
There is too much
light on the SpO
2
sensor.
Move the sensor to a place
with lower level of
ambient light or cover the
sensor to minimize the
ambient light.
SpO
2
No
Pulse
(Mindray,
Nellcor)
No Low
SpO
2
sensor failed
to obtain pulse
signal.
Move the sensor to a site
with better perfusion.
Summary of Contents for VS-900
Page 1: ...VS 900 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 14: ...8 FOR YOUR NOTES ...
Page 26: ...2 6 2 3 4 Bottom View 1 Battery compartment door 2 Hole for installing a support 1 2 ...
Page 42: ...3 10 FOR YOUR NOTES ...
Page 68: ...6 12 FOR YOUR NOTES ...
Page 76: ...7 8 FOR YOUR NOTES ...
Page 78: ...8 2 FOR YOUR NOTES ...
Page 88: ...9 10 FOR YOUR NOTES ...
Page 114: ...12 8 FOR YOUR NOTES ...
Page 139: ...16 11 2 Select Modify Password 3 Enter new password and then select Ok ...
Page 140: ...16 12 FOR YOUR NOTES ...
Page 166: ...B 8 FOR YOUR NOTES ...
Page 191: ......
Page 192: ...P N 046 004883 00 7 0 ...