16-4
Check/Maintenance Item
Recommended Frequency
Preventive Maintenance
Visual inspection
When first installed or after reinstalled.
NIBP Tests
Pressure check
1.
If the user suspects that the measurement is
incorrect.
2.
Following any repairs or replacement of the
module.
3.
At least once a year
Leakage Test
CO
2
Test
Leakage test
Performance test
Calibration
Performance Test
SpO
2
Test
1.
If the user suspects that the measurement is
incorrect.
2.
Following any repairs or replacement of the
module.
3.
At least once every two years for SpO
2
and
Temp, and at least once a year for NIBP and CO
2
.
NIBP test
Pressure check
Leakage test
Temp test
CO
2
test and
calibration
Leakage test
Performance test
Calibration
Nurse Call Relay Performance Test
If user suspects that the nurse call or analog output
does not work well.
Electrical safety tests
Electrical safety tests
At least once every two years, or as required.
Other tests
Power on test
1.
When first installed or after reinstalled.
2.
Following any maintenance or the replacement
of any main unit parts.
Touchscreen Calibration
1.
When the touchscreen appears abnormal.
2.
After the touchscreen is replaced.
Recorder check
Following any repair or replacement of the recorder.
Battery
check
Functionality test
1.
When first installed.
2.
Whenever a battery is replaced.
Performance test
Once a year or if the battery run time reduced
significantly.
Summary of Contents for VS-900
Page 1: ...VS 900 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 14: ...8 FOR YOUR NOTES ...
Page 26: ...2 6 2 3 4 Bottom View 1 Battery compartment door 2 Hole for installing a support 1 2 ...
Page 42: ...3 10 FOR YOUR NOTES ...
Page 68: ...6 12 FOR YOUR NOTES ...
Page 76: ...7 8 FOR YOUR NOTES ...
Page 78: ...8 2 FOR YOUR NOTES ...
Page 88: ...9 10 FOR YOUR NOTES ...
Page 114: ...12 8 FOR YOUR NOTES ...
Page 139: ...16 11 2 Select Modify Password 3 Enter new password and then select Ok ...
Page 140: ...16 12 FOR YOUR NOTES ...
Page 166: ...B 8 FOR YOUR NOTES ...
Page 191: ......
Page 192: ...P N 046 004883 00 7 0 ...