6-8
6.6
Pausing Alarms
You can temporarily disable alarm indicators by pressing and holding
hardkey on
the panel for above 2 seconds.
When alarms are paused,
The alarms pause symbol
and the remaining alarm pause time is displayed
in Alarm Information area.
For physiological alarms, no alarm indication is shown. New physiological alarm
will not be presented.
For technical alarms, alarm sounds are paused, but alarm lamps and alarm
messages remain presented.
The default alarm pause time is 2 minutes. When the alarm pause time expires, or the low
battery alarm occurs, the alarm paused status is automatically deactivated. You can also
cancel the alarm paused status by pressing and holding the
hardkey for more than
2 seconds.
6.7
Switching Off Alarm Sound
When alarm volume is set to 0, the alarm sound is turned off. In audio alarm off status,
The Audio Off symbol
is displayed.
Audio indication of all alarms is suspended.
You can cancel the alarm sound off status by setting alarm volume to a value from 1 to 10.
WARNING
Pausing or switching off alarms may result in a hazard to the patient. Please be
very careful.
6.8
Resetting Alarms
By pressing the
hardkey, you can reset the alarm system to acknowledging the
on-going alarms and enable the alarm system to respond to a subsequent alarm
condition.
For physiological alarms, except the NIBP-related alarms, when the alarm system is reset:
The alarm sound is silenced.
A √ appears before the alarm message, indicating that the alarm is
acknowledged.
Summary of Contents for VS-900
Page 1: ...VS 900 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 14: ...8 FOR YOUR NOTES ...
Page 26: ...2 6 2 3 4 Bottom View 1 Battery compartment door 2 Hole for installing a support 1 2 ...
Page 42: ...3 10 FOR YOUR NOTES ...
Page 68: ...6 12 FOR YOUR NOTES ...
Page 76: ...7 8 FOR YOUR NOTES ...
Page 78: ...8 2 FOR YOUR NOTES ...
Page 88: ...9 10 FOR YOUR NOTES ...
Page 114: ...12 8 FOR YOUR NOTES ...
Page 139: ...16 11 2 Select Modify Password 3 Enter new password and then select Ok ...
Page 140: ...16 12 FOR YOUR NOTES ...
Page 166: ...B 8 FOR YOUR NOTES ...
Page 191: ......
Page 192: ...P N 046 004883 00 7 0 ...