2-11
common mode for monitoring patient vital signs.
NOTE
In Monitor mode, physiological and technical alarms, and prompt messages are
supported.
2.6.2
Spot Check Mode
The spot mode is intended for on-spot measurement in a short period. When spot mode
is ON, the message area on the top of screen will display [
Spot Check
].
To enter Spot Check mode, select [
Main]
→
[
Maintenance>>
]
→
[
User Settings>>
]
→
Enter
required password
→
Set [
Spot Check
] to [
On
].
NOTE
In Spot Check mode, technical alarms and prompt messages are supported, but
no physiological alarms.
Monitor Mode vs. Spot Check Mode
The Monitor mode and Spot Check modes have all the features in common except the
following:
Functions
Monitor Mode
Spot Check Mode
Configure and use Sat-Seconds (Nellcor)
Yes
No
Access [
Alarm Setup
] tab
Yes No
Access [
Continuous Trends
] tab
Yes No
Access [
Graphic Trends
] tab
Yes No
Connect to the central monitoring system
Yes
No
Measure CO
2
Yes
No
2.6.3
Standby Mode
In standby mode, the monitoring for current patient is over, but the monitor still powers
on.
If no parameter is being measured, you can press the power switch to enter Standby
mode. Then a warning pops up. Select [
Yes
] to enter the Standby mode.
When the monitor is powered by a battery, it will automatically enter the Standby mode
when the following conditions are satisfied:
No key operation within 10 minutes.
No unacknowledged alarms.
Summary of Contents for VS-900
Page 1: ...VS 900 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 14: ...8 FOR YOUR NOTES ...
Page 26: ...2 6 2 3 4 Bottom View 1 Battery compartment door 2 Hole for installing a support 1 2 ...
Page 42: ...3 10 FOR YOUR NOTES ...
Page 68: ...6 12 FOR YOUR NOTES ...
Page 76: ...7 8 FOR YOUR NOTES ...
Page 78: ...8 2 FOR YOUR NOTES ...
Page 88: ...9 10 FOR YOUR NOTES ...
Page 114: ...12 8 FOR YOUR NOTES ...
Page 139: ...16 11 2 Select Modify Password 3 Enter new password and then select Ok ...
Page 140: ...16 12 FOR YOUR NOTES ...
Page 166: ...B 8 FOR YOUR NOTES ...
Page 191: ......
Page 192: ...P N 046 004883 00 7 0 ...