12-6
parameter tile has a value. To calculate a score, follow this procedure:
1.
Start monitoring parameters. Their values are automatically obtained.
2.
Manually input the values for the parameters that are not monitored.
NOTE
If the value is measured by the monitor, the value displays in grey. You cannot
manually input the value.
When you are in the scoring screen, if an alarm for the monitoring parameter
occurs, the alarm indicators still can be seen in the monitor display.
The parameter input range is as follows:
Parameter
Range
Pulse Rate
20 bpm -300 bpm
Systolic NIBP
40 mmHg -270 mmHg (Adult)
40 mmHg -200 mmHg (Pediatric)
40 mmHg -135 mmHg (Neonate)
Respiration Rate
0-150 rpm
Temperature 0.1°C
--- 50.0°C (32.1°F --- 122.0°F)
Level of Consciousness
AVPU: Alert, Reacting to Voice, Reacting to Pain,
Unresponsive
GCS: 1-15
Supplemental Oxygen Yes,
No
SpO2 0%
--- 100%
Urine Output
0 -300 mml/h (0 --- 10 ml/h/kg)
Catheter Yes,
No
Blood Sugar
1.0 mg/dl -720.0 mg/dl (0.06 mmol/L -40.00 mmol/L)
Pain Score
0-10
Pain
None, Mild, Moderate, Severe
Inspired O2%
21% -100%
Airway Clear,
Obstruction
Customizable parameter
The input range depends on the decimal point. The decimal
point is customizable in Mindray Clinical Scoring Config Tool.
0 --- 9999 (decimal point as 1)
0.0 --- 999.9 (decimal point as 0.1)
0.00 --- 99.99 (decimal point as 0.01)
Summary of Contents for VS-900
Page 1: ...VS 900 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 14: ...8 FOR YOUR NOTES ...
Page 26: ...2 6 2 3 4 Bottom View 1 Battery compartment door 2 Hole for installing a support 1 2 ...
Page 42: ...3 10 FOR YOUR NOTES ...
Page 68: ...6 12 FOR YOUR NOTES ...
Page 76: ...7 8 FOR YOUR NOTES ...
Page 78: ...8 2 FOR YOUR NOTES ...
Page 88: ...9 10 FOR YOUR NOTES ...
Page 114: ...12 8 FOR YOUR NOTES ...
Page 139: ...16 11 2 Select Modify Password 3 Enter new password and then select Ok ...
Page 140: ...16 12 FOR YOUR NOTES ...
Page 166: ...B 8 FOR YOUR NOTES ...
Page 191: ......
Page 192: ...P N 046 004883 00 7 0 ...