16-7
16.5.2
NIBP Accuracy Test
The NIBP accuracy test is required at least once a year or when you doubt the measured
NIBP.
Tools required:
T-shape connector
Tubing
Balloon pump
Metal Vessel (volume 500±25 ml)
Reference manometer (calibrated with accuracy equal to or better than 0.75
mmHg)
Follow this procedure to perform the accuracy test:
1.
Connect the equipment as shown.
2.
Before inflation, check that the reading of the manometer should be 0. If not, open
the valve of the balloon pump to let the whole airway open to the atmosphere. Close
the valve of the balloon pump after the reading is 0.
3. Select
[
Main
]
→
[
Maintenance >>
]
→
[
User Settings>>
]
→
Enter required password
→
→
[
Module
Maintenance >>
]
→
[
NIBP Accuracy Test
]. The message [
Accuracy
Testing…
] is displayed in the NIBP parameter area.
4.
Check the manometer values and the monitor reading. Both should be 0mmHg.
5.
Raise the pressure in the metal vessel to 50 mmHg with the balloon pump. Then wait
for 10 seconds until the measured values become stable.
6.
Compare the manometer values with the displayed values. The difference between
the manometer and displayed values should be within ± 3 mmHg.
7.
Raise the pressure in the metal vessel to 200 mmHg with the balloon pump. Then
wait for 10 seconds until the measured values become stable. Repeat step 6.
Monitor
Connector for
NIBP cuff
Tubing
Manometer
Metal vessel
Balloon pump
Summary of Contents for VS-900
Page 1: ...VS 900 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 14: ...8 FOR YOUR NOTES ...
Page 26: ...2 6 2 3 4 Bottom View 1 Battery compartment door 2 Hole for installing a support 1 2 ...
Page 42: ...3 10 FOR YOUR NOTES ...
Page 68: ...6 12 FOR YOUR NOTES ...
Page 76: ...7 8 FOR YOUR NOTES ...
Page 78: ...8 2 FOR YOUR NOTES ...
Page 88: ...9 10 FOR YOUR NOTES ...
Page 114: ...12 8 FOR YOUR NOTES ...
Page 139: ...16 11 2 Select Modify Password 3 Enter new password and then select Ok ...
Page 140: ...16 12 FOR YOUR NOTES ...
Page 166: ...B 8 FOR YOUR NOTES ...
Page 191: ......
Page 192: ...P N 046 004883 00 7 0 ...