6-7
Parameter Low alarm Limit
High alarm Limit
Auto-set alarm
limit
Range
Adult/
Pediatric
Neonate
Adult/
Pediatric
Neonate
NIBP-D
(Dia×0.68)
+
6mmHg
(Dia-15) or
20mmHg
,
whichever is
greater
(Dia×0.86)
+
32mmHg
(Dia
+
15) or
80mmHg
,
whichever is
smaller
Adult:
25 to 225 mmHg
Pediatric:
25 to 150 mmHg
Neonate:
20 to 90 mmHg
NIBP-M
(Mean×0.68)
+
8mmHg
(Mean-15) or
35mmHg
,
whichever is
greater
(Mean×0.86)
+
35mmHg
(Mean
+
15) or
95mmHg
,
whichever is
smaller
Adult:
30 to 245 mmHg
Pediatric:
30 to 180 mmHg
Neonate:
25 to 105 mmHg
Temp
(T
-
0.5)°C (T
-
0.5) °C
(T
+
0.5) °C
(T
+
0.5)°C
Same as the
measurement
range
EtCO
2
0 to 32: remains
the same
0 to 32:
remains the
same
0 to 32:
remains the
same
0 to 32:
remains the
same
Same as the
measurement
range
33 to 35: 29
33 to 35: 29
33 to 35: 41
33 to 35: 41
Same as the
measurement
range
36 to 45: (EtCO
2
- 6)
36 to 45:
(EtCO
2
- 6)
36 to 45:
(EtCO
2
+ 6)
36 to 45: (EtCO
2
+ 6)
Same as the
measurement
range
46 to 48: 39
46 to 48: 39
46 to 48: 51
46 to 48: 51
Same as the
measurement
range
>48: remains
the same
>48: remains
the same
>48: remains
the same
>48: remains
the same
Same as the
measurement
range
FiCO
2
None
None
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
measurement
range
Summary of Contents for VS-900
Page 1: ...VS 900 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 14: ...8 FOR YOUR NOTES ...
Page 26: ...2 6 2 3 4 Bottom View 1 Battery compartment door 2 Hole for installing a support 1 2 ...
Page 42: ...3 10 FOR YOUR NOTES ...
Page 68: ...6 12 FOR YOUR NOTES ...
Page 76: ...7 8 FOR YOUR NOTES ...
Page 78: ...8 2 FOR YOUR NOTES ...
Page 88: ...9 10 FOR YOUR NOTES ...
Page 114: ...12 8 FOR YOUR NOTES ...
Page 139: ...16 11 2 Select Modify Password 3 Enter new password and then select Ok ...
Page 140: ...16 12 FOR YOUR NOTES ...
Page 166: ...B 8 FOR YOUR NOTES ...
Page 191: ......
Page 192: ...P N 046 004883 00 7 0 ...