6-5
WARNING
Make sure that the alarm limits settings are appropriate for your patient before
monitoring.
Setting alarm limits to extreme values may cause the alarm system to become
ineffective. For example, High oxygen levels may predispose a premature infant
to retrolental fibroplasia. If this is a consideration do NOT set the high alarm
limit to 100%, which is equivalent to switching the alarm off.
6.5
Selecting Alarm Properties
6.5.1
Changing the Alarm Volume
Select [
Main
]
→
[
General Setup >>
]
→
[
Alarm Volume
].
The alarm volume range is between X to 10. X is the minimum volume, which depends on
the setting of minimum alarm volume (Refer to
6.5.2 Setting the Minimum Alarm
Volume
), and 10 is the maximum volume.
When alarm volume is set to 0, the alarm sound is turned off and a
symbol appears
on the screen.
6.5.2
Setting the Minimum Alarm Volume
1. Select
[
Main
]
→
[
Maintenance >>
]
→
[
User Settings >>
]
→
Enter the required
password
→
Select [
Ok
] to access [
User Settings
] menu.
2. Select
[
Alarm Setup >>
] to access [
Alarm Setup
] menu.
3. Select
[
Minimum Alarm Volume
] and then select the appropriate settings.
The minimum alarm volume refers to the minimum value you can set for the alarm
volume, which is not affected by user or factory default configurations.
6.5.3
Setting the Interval between Alarm Sounds
1. Select
[
Main
]
→
[
Maintenance >>
]
→
[
User Settings >>
]
→
Enter the required
password
→
Select [
Ok
] to access [
User Settings
] menu.
2. Select
[
Alarm Setup >>
] to access [
Alarm Setup
] menu.
3. Select
[
High Alarm Interval (s)
], [
Med Alarm Interval (s)
] and [
Low Alarm Interval
(s)
] in turn and then select the appropriate settings.
Summary of Contents for VS-900
Page 1: ...VS 900 Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 14: ...8 FOR YOUR NOTES ...
Page 26: ...2 6 2 3 4 Bottom View 1 Battery compartment door 2 Hole for installing a support 1 2 ...
Page 42: ...3 10 FOR YOUR NOTES ...
Page 68: ...6 12 FOR YOUR NOTES ...
Page 76: ...7 8 FOR YOUR NOTES ...
Page 78: ...8 2 FOR YOUR NOTES ...
Page 88: ...9 10 FOR YOUR NOTES ...
Page 114: ...12 8 FOR YOUR NOTES ...
Page 139: ...16 11 2 Select Modify Password 3 Enter new password and then select Ok ...
Page 140: ...16 12 FOR YOUR NOTES ...
Page 166: ...B 8 FOR YOUR NOTES ...
Page 191: ......
Page 192: ...P N 046 004883 00 7 0 ...