6 Acquiring Images
Basic User Manual
87
Figure 6-7 PDI-Mode Imaging Screen
3.
Adjust color flow ROI.
−
Move the trackball to position color flow ROI.
−
Press the confirm key on the control panel to adjust the size of color flow ROI.
−
Press the confirm key again to reposition color flow ROI.
4.
Optimize the PDI-mode image. For details, refer to Section 6.3.4 Optimizing CFM/
PDI/TDI Mode Images.
5.
Tap
PDI
again to exit the screen.
6
�
3
�
3 TDI Mode
NOTE:
TDI imaging is only applied to cardiac applications by using the phased array probes.
TDI (Tissue Doppler Imaging) is a color flow imaging technique which detects the low
frequency signal reflected from the cardiac muscle. TDI provides the flow information of
velocity and direction for cardiac movement.
TDI uses the low velocity and the high amplitude adjusted from the wall filter to create a
color-coded tissue imaging.
Perform the following steps to acquire TDI-mode images.
1.
Optimize a B-mode image.
2.
Press the
CFM
key on the control panel, and then tap
TDI
on the touch screen to
enter the TDI mode.
2D Imaging
Color Flow ROI
PWR 70
FRQ 2.2
15
1.0
86
255
1/50
C
PRF
WF
GN
C/P
10
PWR 70
FRQ 3-4.8
D
16.5cm
5
47
3/1
255
3/30
0
B
FPS
D/G
GN
I/P
Summary of Contents for P25 EXP
Page 1: ...User Manual P25 EXP Ultrasound System Version 1 4...
Page 5: ......
Page 12: ...Contents VIII Basic User Manual D 3 Disinfectant 200 AppendixEAcousticOutputData 201...
Page 48: ...This page is intentionally left blank...
Page 168: ...This page is intentionally left blank...
Page 176: ...This page is intentionally left blank...
Page 200: ...This page is intentionally left blank...
Page 213: ...201 Appendix E Acoustic Output Data Please consult the accompanying CD...