6 Acquiring Images
90
Basic User Manual
Frequency
Frequency optimizes the penetrability and the sensitivity of the color flow.
To adjust the frequency:
●
Tap the left part of
Frequency
on the touch screen to decrease the value.
●
Tap the right part of
Frequency
on the touch screen to increase the value.
Wall Filter
Wall filter is used to filter the low frequency noise from tissues, such as the vascular wall.
To adjust the wall filter:
●
Tap the left part of
WF
on the touch screen to decrease the value.
●
Tap the right part of
WF
on the touch screen to increase the value.
Pulse Repetition Frequency
Pulse repetition frequency is used to adjust the velocity scale.
To adjust the pulse repetition frequency:
●
Tap the left part of
PRF
on the touch screen to decrease the value.
●
Tap the right part of
PRF
on the touch screen to increase the value.
Line Density
Line density refers to the amount of ultrasound beam that makes up the image. Increasing
the value of line density improves the resolution and decreases frame rate. Therefore, you
have to balance the frame rate and the image quality when adjusting the line density.
A lower line density is useful in fetal heart beat, adult cardiac applications and in clinical
radiology applications. A higher line density is useful in obtaining a very high resolution,
such as for the thyroid or testicles.
To adjust the line density:
●
Tap the left part of
Line Density
on the touch screen to decrease the value.
●
Tap the right part of
Line Density
on the touch screen to increase the value.
Baseline
Baseline represents the position of zero velocity or frequency. The velocity range in one
direction can be increased or decreased by adjusting the baseline.
Baseline is used to eliminate the alias in the color flow imaging, and display higher
velocities without reversal of colors.
To adjust the baseline:
Summary of Contents for P25 EXP
Page 1: ...User Manual P25 EXP Ultrasound System Version 1 4...
Page 5: ......
Page 12: ...Contents VIII Basic User Manual D 3 Disinfectant 200 AppendixEAcousticOutputData 201...
Page 48: ...This page is intentionally left blank...
Page 168: ...This page is intentionally left blank...
Page 176: ...This page is intentionally left blank...
Page 200: ...This page is intentionally left blank...
Page 213: ...201 Appendix E Acoustic Output Data Please consult the accompanying CD...