Basic User Manual
8 Contrast Imaging
!
Contrast imaging is only used in conjunction with commercial contrast agents legally
approved by laws and local regulations. Do not use the contrast agent not got any legal
permits or official registrations. The manufacturer would not guarantee the safety and
reliability for any contrast agents.
The following description uses a liver examination performed with the C353 probe as an
example.
8
�
1 Entering Contrast Imaging
Perform the following steps to enter the contrast imaging.
1.
Select C353 and Abdomen as the desired probe and exam type, the system
automatically enters the real-time B mode.
2.
Position the target tissue in the B+Color mode, and observe the blood flow.
3.
Return to the B mode, tap
Contrast
on the touch screen to enter the contrast imaging.
As Figure 8-1 shows, a real-time contrast image is displayed on the left and a B-mode
in real-time mode is displayed on the right.
Figure 8-1 Contrast Imaging Screen
−
Tap
Single View
to enter the single display.
−
Tap
Dual View
to enter the dual split display.
−
Tap
Exit
to exit the imaging screen.
T1: 00:20/00:15
T2: 00:15/00:10
15
PWR 4
10
110
5/30
GN
C/P
PWR 4
5
FRQ 3-5.0
D
16.5cm
0
10
200/5
255
3/30
FPS
D/G
GN
I/P
Summary of Contents for P25 EXP
Page 1: ...User Manual P25 EXP Ultrasound System Version 1 4...
Page 5: ......
Page 12: ...Contents VIII Basic User Manual D 3 Disinfectant 200 AppendixEAcousticOutputData 201...
Page 48: ...This page is intentionally left blank...
Page 168: ...This page is intentionally left blank...
Page 176: ...This page is intentionally left blank...
Page 200: ...This page is intentionally left blank...
Page 213: ...201 Appendix E Acoustic Output Data Please consult the accompanying CD...