6 Acquiring Images
86
Basic User Manual
Figure 6-6 CFM-Mode Imaging Screen
3.
Adjust color flow ROI.
−
Move the trackball to position color flow ROI.
−
Press the confirm key on the control panel to adjust the size of color flow ROI.
−
Press the confirm key again to reposition color flow ROI.
4.
Optimize the CFM-mode image. For details, refer to Section 6.3.4 Optimizing CFM/
PDI/TDI Mode Images.
5.
Press the
CFM
key again to exit the screen.
6
�
3
�
2 PDI Mode
PDI (Power Doppler Imaging) is a color flow imaging technology which adds the flow
signal in the CFM-mode image. PDI uses the number and amplitude of red blood cells
going through in the flow to create the color-coded imaging. A flow with slow velocity
and small rate can be displayed in PDI. Therefore, a flow with a higher sensitivity can be
detected without overlaying any flows with high velocity.
The direction of flow, the flow velocity and flow properties cannot be reflected in the PDI-
Mode image.
Perform the following steps to acquire PDI-mode images:
1.
Optimize a B-mode image.
2.
Press the
CFM
key on the control panel, and then tap
PDI
on the touch screen to enter
the PDI mode.
2D Imaging
Color Flow ROI
PWR 70
FRQ 2.2
15
cm/s
1.0
86
255
1/50
-9
10
FRQ 3-4.8
D
16.5cm
C
PRF
WF
GN
C/P
5
PWR 70
9
0
47
3/1
255
3/30
B
FPS
D/G
GN
I/P
Summary of Contents for P25 EXP
Page 1: ...User Manual P25 EXP Ultrasound System Version 1 4...
Page 5: ......
Page 12: ...Contents VIII Basic User Manual D 3 Disinfectant 200 AppendixEAcousticOutputData 201...
Page 48: ...This page is intentionally left blank...
Page 168: ...This page is intentionally left blank...
Page 176: ...This page is intentionally left blank...
Page 200: ...This page is intentionally left blank...
Page 213: ...201 Appendix E Acoustic Output Data Please consult the accompanying CD...