![MediSono P25 EXP User Manual Download Page 161](http://html1.mh-extra.com/html/medisono/p25-exp/p25-exp_user-manual_1762413161.webp)
11 Working with Images
Basic User Manual
149
The ECG module provided by this ultrasound system is for collecting and displaying 3-lead
ECG trace.
!
●
This ECG module is not suitable for intracardiac use or direct cardiac contact.
●
This ECG module provides 3-channel ECG signals and cannot be used for diagnosis
and monitoring.
●
To avoid electrical shock, ensure the ECG cable is intact and be correctly connected.
●
Conductive parts of electrodes and connectors for ECG should not come into contact
with other conductive parts including earth/grounding.
●
Power off the system if the patient is found to use a pacemaker or implantable
cardioverter-defibrillators, because the system may interfere with these devices.
11
�
6
�
1 Basic Procedures of ECG Operation
Perform the following steps:
1.
Power off the system and connect the ECG cable to the specific port of the system.
2.
Power on the system and attach the ECG electrodes on the patient’s body (as shown in
the following figure).
3.
Make the relevant settings of the ECG module. For details, refer to Section 11.6.2
Optimizing Parameters.
4.
Freeze the images and review them. For details, refer to Section 11.6.3 Reviewing
ECG.
5.
Exit the ECG mode, and remove ECG electrodes from the patient.
Summary of Contents for P25 EXP
Page 1: ...User Manual P25 EXP Ultrasound System Version 1 4...
Page 5: ......
Page 12: ...Contents VIII Basic User Manual D 3 Disinfectant 200 AppendixEAcousticOutputData 201...
Page 48: ...This page is intentionally left blank...
Page 168: ...This page is intentionally left blank...
Page 176: ...This page is intentionally left blank...
Page 200: ...This page is intentionally left blank...
Page 213: ...201 Appendix E Acoustic Output Data Please consult the accompanying CD...