6 Acquiring Images
Basic User Manual
99
−
Adjust the size of the sample volume gate by using the trackball.
−
Rotate the
Angle
knob on the control panel to adjust the angle of the flow cursor.
5.
Press the
Update
key on the control panel to activate the PW mode.
The PW spectrum is displayed at the lower part of the screen after being activated, as
shown in Figure 6-15.
Figure 6-15 PW-Mode Imaging Screen
−
X-axis is the time scale.
−
Y-axis is Doppler frequency scale, including positive and negative indicators.
Rotate the
Audio
knob on the control panel to adjust the audio volume.
NOTE:
Once the PW-mode is activated, you can move the trackball to stop the Doppler
trace and adjust the sample volume gate. The system continues tracing if the
trackball is idle for more than 0.5s.
6.
Optimize the PW-mode image. For details, refer to Section 6.5.3 Optimizing Spectral
Doppler Images.
If needed, press the
Update
key on the control panel to enter the inactivated B+PW
mode 2, the following screen appears.
X-axis: Time
-50
Y-axis: Frequency
cm/s
50
Sample Volume Gate
2D Imaging
5
Spectral Doppler Line
Flow Cursor
0
Summary of Contents for P25 EXP
Page 1: ...User Manual P25 EXP Ultrasound System Version 1 4...
Page 5: ......
Page 12: ...Contents VIII Basic User Manual D 3 Disinfectant 200 AppendixEAcousticOutputData 201...
Page 48: ...This page is intentionally left blank...
Page 168: ...This page is intentionally left blank...
Page 176: ...This page is intentionally left blank...
Page 200: ...This page is intentionally left blank...
Page 213: ...201 Appendix E Acoustic Output Data Please consult the accompanying CD...