14 Probes and Biopsy
Basic User Manual
169
3.
Perform a scan.
14
�
1
�
3 Cleaning the Probe
!
●
To avoid electrical shock, disconnect the probe from the system before cleaning.
●
To avoid potential disease transmission, you should wear sterile medical gloves and
protective goggles while cleaning.
●
Do not use cleaner that contains sodium sulfate and/or sodium hypochlorite to clean
the probe. Otherwise, the probe may be damaged.
●
To avoid probe damages, do not bump the probe on hard surfaces.
●
Use only the cleaner recommended by the manufacturer to clean the probe. Otherwise,
the probe may be damaged.
You should clean the probe before and after each use. Probe cleaning is an essential step of
the disinfection and sterilization.
Perform the following steps to clean the probe.
1.
Remove the coupling gel from the probe surface after exam.
2.
Disconnect the probe from the system and remove the probe sheath and biopsy
bracket from the probe.
3.
Use a lint-free soft cloth dampened with mild soapy water or the recommended
cleaner (see D.2 Cleaner) to wipe the probe.
If there are dried stains on the surface or in the grooves of the probe, remove them by
a brush dampened with an enzymatic clearer.
4.
Rinse the surfaces of the probe with the fresh running water to completely remove any
residual cleaner.
5.
Dry the probe with a lint-free soft dry cloth.
Summary of Contents for P25 EXP
Page 1: ...User Manual P25 EXP Ultrasound System Version 1 4...
Page 5: ......
Page 12: ...Contents VIII Basic User Manual D 3 Disinfectant 200 AppendixEAcousticOutputData 201...
Page 48: ...This page is intentionally left blank...
Page 168: ...This page is intentionally left blank...
Page 176: ...This page is intentionally left blank...
Page 200: ...This page is intentionally left blank...
Page 213: ...201 Appendix E Acoustic Output Data Please consult the accompanying CD...