6 Acquiring Images
Basic User Manual
101
0
5
50
cm/s
Y-axis: Frequency
-50
X-axis: Time
−
The spectral Doppler line and the sample volume gate are used to locate the
qualitative analysis on the image.
−
The flow cursor needs to be adjusted parallel to the flow when measuring the
velocity.
3.
Adjust the position and angle of the spectral Doppler line.
−
Position the sample volume gate on the spectral Doppler line by moving the
trackball upwards or downwards.
−
Adjust the angle of the spectral Doppler line by moving the trackball to the left or
the right.
4.
Rotate the
Angle
knob on the control panel to adjust the angle of the flow cursor.
5.
Press the
Update
key on the control panel to activate the CW mode.
The CW spectrum is displayed at the lower part of the screen after being activated, as
shown in Figure 6-18.
Figure 6-18 CW-Mode Imaging Screen
−
X-axis is the time scale.
−
Y-axis is Doppler frequency scale, including positive and negative indicators.
Rotate the
Audio
knob on the control panel to adjust the audio volume.
NOTE:
Once the CW-mode is activated, you can move the trackball to stop the Doppler
trace and adjust the sample volume gate. The system continues tracing if the
trackball is idle for more than 0.5s.
Summary of Contents for P25 EXP
Page 1: ...User Manual P25 EXP Ultrasound System Version 1 4...
Page 5: ......
Page 12: ...Contents VIII Basic User Manual D 3 Disinfectant 200 AppendixEAcousticOutputData 201...
Page 48: ...This page is intentionally left blank...
Page 168: ...This page is intentionally left blank...
Page 176: ...This page is intentionally left blank...
Page 200: ...This page is intentionally left blank...
Page 213: ...201 Appendix E Acoustic Output Data Please consult the accompanying CD...