15 System Maintenance
186
Basic User Manual
●
Connectivity of interfaces
Image Recording
●
Images in each mode
●
Image recording using the standard probe
15
�
3 Troubleshooting
A1: Cannot start the system.
Q1: Check whether the system is plugged in, the main power switch is located to
position and the fuse is intact. If the fuse is blown, replace it as described in Section 15.4
Replacing the Fuse.
A2: No images are displayed on the LCD monitor.
Q2: Check whether the power supply cable and video cable are securely connected.
A3: The LCD monitor displays abnormal.
Q3: Check whether you select the appropriate probe and a compatible exam type, or
optimize imaging parameters correctly. For details, refer to Chapter 6 Acquiring Images.
A4: The measurement cannot be performed after pressing the
Calc
key on the control
panel.
Q4: Ensure you pressing the
Calc
key in the frozen mode.
If any of the above problems still exist, please contact the Customer Service Department of
the manufacturer for help.
15
�
4 Replacing the Fuse
!
Power off the system and disconnect the power cable from the socket before replacing the
fuse.
Two fuses with specifications of 50CT-T5AH 250V are provided by the
manufacturer. You can also purchase the fuse with the same specifications
for the fuse replacement.
You can replace the fuse under the instructions of the service personnel if
the fuse is blown.
If the fuse is blown again after a replacement, please stop using the
system immediately and contact customer service.
Fuse
50CT-T5AH 250V
T1 T2
Summary of Contents for P25 EXP
Page 1: ...User Manual P25 EXP Ultrasound System Version 1 4...
Page 5: ......
Page 12: ...Contents VIII Basic User Manual D 3 Disinfectant 200 AppendixEAcousticOutputData 201...
Page 48: ...This page is intentionally left blank...
Page 168: ...This page is intentionally left blank...
Page 176: ...This page is intentionally left blank...
Page 200: ...This page is intentionally left blank...
Page 213: ...201 Appendix E Acoustic Output Data Please consult the accompanying CD...