14 Probes and Biopsy
Basic User Manual
173
Table 14-3 Recommended Method for Sterilization
Sterilant
Manufacturer Active Ingredients Level for Active
Ingredient
Contact Type Contact Period
Cidex™ Activated
D i a l d e h y d e
Solution
J&J
Glutaraldehyde
2.4%
Soak
10 hours
14
�
1
�
5 Disinfecting and Sterilizing the Probe Cable
!
To avoid potential disease transmission, you should wear sterile medical gloves and
protective goggles during disinfection or serialization.
Do not immerse the probe cable in alcohol or isopropyl alcohol at any concentration for a
long period of time. Avoid any contact with iodine or phenols. Otherwise, it will accelerate
the aging of the cable and decrease the service time of the probe, or even result in cable
damages.
Disinfection
Perform the following steps to disinfect the probe cable.
1.
Use a lint-free soft cloth dampened with mild soapy water to wipe the probe cable.
2.
Rinse the cable with fresh running water to completely remove the residual cleaner.
3.
Perform the medium-level disinfection to the probe cable as shown in Table14-4.
4.
Rinse the surfaces of the probe cable with fresh running water to completely remove
the residual disinfectant.
5.
Dry the probe cable with a lint-free soft dry cloth.
Table 14-4 Recommended Method for Medium-level Disinfection
Disinfectant Manufacturer Active Ingredients
Level for Active
Ingredient
Contact Type
Contact Period
T-spray II
Pharm. Inc.
Quat.Ammonia
/
Spray/Wipe
<10 minutes
T-spray
Pharm. Inc.
Quat.Ammonia
/
Spray/Wipe
<10 minutes
Sterilization
If the probe cable contacts with any blood or body fluid you should sterilize it. For details,
refer to the information about probe sterilization.
Summary of Contents for P25 EXP
Page 1: ...User Manual P25 EXP Ultrasound System Version 1 4...
Page 5: ......
Page 12: ...Contents VIII Basic User Manual D 3 Disinfectant 200 AppendixEAcousticOutputData 201...
Page 48: ...This page is intentionally left blank...
Page 168: ...This page is intentionally left blank...
Page 176: ...This page is intentionally left blank...
Page 200: ...This page is intentionally left blank...
Page 213: ...201 Appendix E Acoustic Output Data Please consult the accompanying CD...