1 Safety
Basic User Manual
3
●
The video printer should be connected to the specific interface using the cable
provided by the manufacturer; Otherwise, electrical shock may occur.
●
In the environment that a patient is 1.8 meters (6 feet) around, connect peripherals
to the auxiliary power outlet which is capable of isolation protection, or power the
peripherals by auxiliary output cable or isolation transformer complied with EN/IEC
60601-1 or the power input of the same safety level.
●
Parts of non-medical electrical equipment in the patient environment that, after
removal of covers, connectors, etc., without the use of a tool, may be contacted during
routine maintenance, calibration, etc. You should not touch the referred parts and the
patient simultaneously.
●
Use only the probes provided by the manufacturer. Otherwise, the system will not
perform, and an accident such as a fire may result in the worst case.
●
Only the peripherals and accessories provided or recommended by the manufacturer
can be used. Using other devices or accessories may degrade the system’s performance
and even cause an electrical shock.
●
Do not pour any fluid onto the system’s surfaces, as fluid seepage into the electrical
circuitry may cause excessive electrical currenet leakage or system failure. If any
water is carelessly spilled onto the system, immediately stop using the system and
contact Service Representative immediately.
●
If this system is transported to the operating environment with a great temperature
change, leave it for approx. 4 hours before powering it on. Ensure temperature and
humidity inside and around this system are equivalent before an operation.
●
Do not use this system around a strong electric field, a strong electromagnetic field
or the devices which generate radio waves, such as a radio, cellular telephones, or
transceivers. Using the system in an improper environment may result in malfunction
or damage.
●
Do not use this system at the same time with other equipment such as electric knife,
high-frequency therapy equipment and defibrillator. Otherwise, electrical shock may
occur.
●
The AC power connector plug for the system is a three-prong grounded plug and
should never be adapted to any two-prong outlet or by using an adapter. The AC
power connector plug should be plugged into a hospital-grade power outlet.
●
Select the conformed multi-socket outlet with protective grounding and ensure that its
maximum output power exceeds the requirement of this system.
Summary of Contents for P25 EXP
Page 1: ...User Manual P25 EXP Ultrasound System Version 1 4...
Page 5: ......
Page 12: ...Contents VIII Basic User Manual D 3 Disinfectant 200 AppendixEAcousticOutputData 201...
Page 48: ...This page is intentionally left blank...
Page 168: ...This page is intentionally left blank...
Page 176: ...This page is intentionally left blank...
Page 200: ...This page is intentionally left blank...
Page 213: ...201 Appendix E Acoustic Output Data Please consult the accompanying CD...