11 Working with Images
Basic User Manual
143
NOTE:
•
Panoramic imaging is only available performed with linear or curved probes.
•
Apply an adequate amount of coupling gel along the intended path of the scan.
•
Do not shake, rotate or tilt the probe during the scan.
•
Ensure the probe remains to be in contact with the skin perpendicularly during the
scan.
•
Do not change the scan direction or move the probe backwards and forwards.
•
When the scan depth increases, generally it is generally required to decrease the
scan speed.
Real-Time Panoramic Imaging
Perform the following steps to apply panoramic imaging.
1.
Optimize the B-mode image.
2.
Tap
Pa
on the touch screen to enter the inactivated Panoramic Imaging
mode.
You can press the
Freeze
key on the control panel to exit and return to the B mode.
3.
Press the
Update
key on the control panel to enter the real-time panoramic imaging.
4.
Perform the scan.
A real-time stitched image is generated, on which a probe speed box and a hint are
displayed. A blue speed box indicates a slower scanning speed, a green speed box
indicates a normal scanning speed and a red speed box indicates a faster scanning
speed.
5.
Press the
Freeze
key to display the panorama.
Figure 11-4 Real-Time Panoramic Imaging Screen
B
FPS
D/G
GN
I/P
PWR
FRQ
D
Summary of Contents for P25 EXP
Page 1: ...User Manual P25 EXP Ultrasound System Version 1 4...
Page 5: ......
Page 12: ...Contents VIII Basic User Manual D 3 Disinfectant 200 AppendixEAcousticOutputData 201...
Page 48: ...This page is intentionally left blank...
Page 168: ...This page is intentionally left blank...
Page 176: ...This page is intentionally left blank...
Page 200: ...This page is intentionally left blank...
Page 213: ...201 Appendix E Acoustic Output Data Please consult the accompanying CD...