6 Acquiring Images
82
Basic User Manual
Chroma
Chroma is used to colorize the gray scale image to enhance the discrimination capability.
To adjust the chroma:
Tap the left or right part of
Chroma
on the touch screen to make a setting.
Compound Imaging
Compound imaging is used to acquire a series of overlapping image frames from
substantially differing spatial directions and combining these images to reduce speckle and
improve contrast resolution.
Tap
Compound
on the touch screen to enable or disable the compound imaging function.
Line Density
Line density refers to the amount of ultrasound beam that makes up the image. Increasing
the value of line density improves the resolution and decreases frame rate. Therefore, you
have to balance the frame rate and the image quality when adjusting the line density.
A higher line density is useful in obtaining a very high resolution, such as the thyroid or
testicles. A lower line density is useful when scanning fetal heart beat and adult cardiac
applications.
To adjust the line density
●
Tap the left part of
Line Density
on the touch screen to decrease the value.
●
Tap the right part of
Line Density
on the touch screen to increase the value.
Persistence
Persistence is used to average consecutive frames to provide a smoother appearance with
less noise. Use lower persistence values for fast-moving organs or tissues and higher
persistence values for a smoother appearance.
To adjust the persistence:
●
Tap the left part of
Persist
on the touch screen to decrease the value.
●
Tap the right part of
Persist
on the touch screen to increase the value.
Dynamic Range
Dynamic range increases the adjustable range of contrast by enhancing the intensities of
parts in the grayscale. It optimizes the ultrasound image texture.
To adjust the dynamic range:
●
Tap the left part of
DYN
on the touch screen to decrease the value.
●
Tap the right part of
DYN
on the touch screen to increase the value.
Summary of Contents for P25 EXP
Page 1: ...User Manual P25 EXP Ultrasound System Version 1 4...
Page 5: ......
Page 12: ...Contents VIII Basic User Manual D 3 Disinfectant 200 AppendixEAcousticOutputData 201...
Page 48: ...This page is intentionally left blank...
Page 168: ...This page is intentionally left blank...
Page 176: ...This page is intentionally left blank...
Page 200: ...This page is intentionally left blank...
Page 213: ...201 Appendix E Acoustic Output Data Please consult the accompanying CD...