189
Appendix A Specifications
Comply with
EN 60601-1 (IEC 60601-1), Medical electrical equipment Part 1: General
requirements for basic safety and essential performance, Class I, BF, continuous
operation
EN 60601-2-37:2008 (IEC 60601-2-37:2007), Medical Electrical Equipment Part
2-37: Particular Requirements for the Basic Safety and Essential Performance of
Ultrasonic Medical Diagnostic and Monitoring Equipment
EN 60601-1-2:2007 (IEC 60601-1-2: 2007), Class A, Medical electrical
equipment- Part 1-2: General requirements for basic safety and essential
performance - Collateral standard: Electromagnetic compatibility - Requirements
and tests
Classifications
Type of protection against
electrical shock
Class I
Degree of protection
against electrical shock
Type-BF applied part
Installation Type
Mobile equipment
Degrees of protection
against harmful liquid
System is IPX0
Probe (from the acoustic window to the junction
line) is IPX7
Footswitch is IPX4
According to the degree
of safety of application
The equipment is not suitable for use in the
presence of a flammable anesthetic mixture with
air, oxygen or nitrous oxide.
Environmental
Requirement
Operations
Storage and Transportation
Ambient Temperature
10°C~40°C
-20°C~+55°C
Relative Humidity
30%~75%
(no condensation)
20%~90%
(no condensation)
Atmospheric Pressure
700hPa~1060hPa
700hPa~1060hPa
Power Supply
100-240V~, 2.7-1.1A, 50/60Hz
Summary of Contents for P25 EXP
Page 1: ...User Manual P25 EXP Ultrasound System Version 1 4...
Page 5: ......
Page 12: ...Contents VIII Basic User Manual D 3 Disinfectant 200 AppendixEAcousticOutputData 201...
Page 48: ...This page is intentionally left blank...
Page 168: ...This page is intentionally left blank...
Page 176: ...This page is intentionally left blank...
Page 200: ...This page is intentionally left blank...
Page 213: ...201 Appendix E Acoustic Output Data Please consult the accompanying CD...